Numerical analysis of an influence of the artificial valve type on the blood flow inside the ventricular assist device
Przemyslaw Klosinski
, Piotr Reorowicz, Damian Obidowski, Krzysztof Jozwik
Technical University of Lodz, Institute of Turbomachinery,,Medical Apparatus Division, Lodz, Poland.
DOI:
https://doi.org/10.7494/cmms.2011.1.0336
Abstract:
A Ventricular Assist Device (VAD) is used in case of severe heart illnesses when the natural heart supplies the body with an insufficient volume of blood. Any damage or improper functioning of the VAD can result in the patient’s death. This implies the constant need to improve the design of VADs and artificial valves which are crucial parts of the device. The authors used the latest Computer Aided Design and Computational Fluid Dynamics software to analyze the flow in the pneumatic Ventricular Assist Device designed at the Foundation of Cardiac Surgery Development and equipped with two different types of valves. In the study, a single-disc mechanical artificial heart valve based on the invention of Prof. J.J. Moll, modified at the Institute of Turbomachinery, TU Lodz, was compared with a three-leaflet polyurethane artificial heart. A comparison was made on the basis of the flow visualization inside the VAD chamber and the size of stagnation regions where the flowing blood may coagulate. An angular position of the disc valve was determined on the basis of the previous studies. A steady state simulation was performed on the assumption that walls of the assist device, adapters and valves were rigid. Dynamic viscosity of blood was defined on the basis on the Non-Newtonian Power Law. Simulations were preformed for systole and diastole conditions. The Ansys CFX v12 code was used to perform preprocessing, solving and postprocessing stages. Deformations of the three-leaflet polyurethane valve were obtained in SolidWorks 2009 and imported to Ansys ICEM v.12. On the basis of the preformed analysis, it has been proven that the disc mechanical heart valve generates better flow conditions inside the heart chamber, especially if a risk of coagulation is concerned. Moreover, the flow observed inside the chamber when the disc valve was used is more homogenous and a single swirl occurring in the central part enables good washing of the connection of the diaphragm and chamber regions.
Cite as:
Klosinski, P., Reorowicz, P., Obidowski, D., & Jozwik, K. (2011). Numerical analysis of an influence of the artificial valve type on the blood flow inside the ventricular assist device. Computer Methods in Materials Science, 11(1), 209 – 214. https://doi.org/10.7494/cmms.2011.1.0336
Article (PDF):

Keywords:
Ventricular assist device, Artificial heart, Artificial heart valves, Numerical fluid mechanics
References:
Bujok, W., Obidowski, D., Kapis, A., Kustosz, R., Jóźwik, K, Reorowicz, P., Kłosiński, P., 2010, Influence of the type of artificial valve onto the flow inside of pediatric ventricular assist device, Journal of Artificial Organs, 33 (7), 471.
De Hart, J., Baaijens, F.P.T., Peters, G.W.M., Schreurs, P.J.G., 2003, A computational fluid-structure interaction analysis of a fiber-reinforced stentless aortic valve, Journal of Biomechanics, 36, 699-712.
Gijsen, F.J.H., Allanie, E., van de Vosse, F.N., Janssen, J.D., 1999, The Influence of the Non-Newtonian Properties of Blood on the Flow in Large Arteries: Unsteady flow in a 90° Curved Tube, Journal of Biomechanics, 32, 705-713.
Gijsen, F.J.H., van de Vosse, F.N., Janssen, J.D., 1999, The Influence of the Non-Newtonian Properties of Blood on the Flow in Large Arteries: Steady Flow in a Carotid Bifurcation Model, Journal of Biomechanics, 32, 601-608.
Johnston, B., Johnson, P., Corney, S., Kilpatrick, D., 2004, Non-Newtonian Blood Flow in Human Right Coronary Arteries: Steady State Simulation, Journal of Biomechanics, 37, 709-720.
Jóźwik, K, Obidowski, D., Kłosiński, P., 2009, Modifications of an Artificial Ventricle Assisting Heart Operation on the Basis of Numerical Methods, Turbomachinery, 135, 61-68.
Jóźwik, K, Obidowski, D., 2010, Numerical Simulations of the Blood Flow through Vertebral Arteries Journal of Biomechanics, 43, 177-185.
Jóźwik, K, Obidowski, D., Reorowicz, P., Kłosiński, P., 2009, Selection of a Heart Valve Prosthesis for the Artificial Heart, The International Journal of Artificial Organs, Special issue dedicated to the ABSTRACTS from the XXXVI ESAO Congress, 2-5 September 2009, Compiegne, France, 25, 32 (7), 429.
Obidowski D., Kłosiński P., Reorowicz P., Jóźwik K, 2010, Influence of an Artificial Valve Type on the Flow in the Ventricular Assist Device, MEDICON 2010, IFMBE Proceedings, 29, 410-413.
Obidowski D., Mysior M., Jóźwik K, 2008, Comparison of Ultrasonic Measurement and Numerical Simulation Results of the Flow through Vertebral Arteries. In: 4th European Conf. of the International Federation for Medical and Biological Engineering, Antwerp, Vol. 22. 286-292.
Walburn, F.J., Schneck, D.J., 1976, A constitutive equation for whole human blood, Biorheology, 31, 201-218.